Instruction manual for the A&D Medical UA-1030T / UA-1030TCN Premier Talking Blood Pressure Monitor — a clinically validated, oscillometric home blood pressure and pulse monitor with TriCheck averaging mode.
Instruction manual for the A&D Medical UA-1030T / UA-1030TCN Premier Talking Blood Pressure Monitor — a clinically validated, oscillometric home blood pressure and pulse monitor with a talking display and TriCheck™ averaging mode.
This blood pressure monitor was clinically validated by an independent organization; measurements are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by ANSI/AAMI SP10 for electronic sphygmomanometers.
The UA-1030T is designed for home use by those 18 years and older to monitor blood pressure (systolic and diastolic) and pulse rate. It is not designed for ambulatory use. The device is based on the oscillometric method — a noninvasive blood pressure determination that examines the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat. The cuff is inflated until the artery is fully blocked; measurements are taken during both inflation and deflation for a faster, more comfortable reading.
Install 4 type AA (1.5V) batteries (alkaline recommended), or use the AC Adapter, and attach the cuff to the monitor before use.
Using the correct cuff size is important for an accurate reading. A cuff that is too large will produce a reading lower than the correct blood pressure; a cuff that is too small will produce a reading higher than the correct blood pressure. With your arm hanging at your side, measure the circumference of your upper arm at the midpoint between the shoulder and elbow. The SlimFit™ cuff accommodates a wide range of arm sizes, from 9–14.6 in. (23–37 cm). If your arm circumference is outside that range, contact A&D Medical (1-888-726-9966) or Auto Control in Canada (1-800-461-0991) for alternate cuff sizes.
After setting the date/time, the next mode is pressure setting. Default is AUTO (automatic). For more rapid readings, you can adjust the default to about 30 mmHg or more above your expected systolic pressure using the arrow buttons, then press the clock button to confirm. The next setting is TriCheck™ Mode.
An optional feature that automatically takes three consecutive readings one minute apart and provides the average, complying with the American Heart Association's recommended measuring method. Default is off; use the arrow buttons to turn on/off, then press the clock button to confirm and end the settings session. To adjust any values, push the clock button and begin again.
The TriCheck symbol displays during the measurement sequence. To cancel, press START (no data is stored). If the cuff fit error symbol appears during the first measurement, cancel, reapply the cuff correctly, and start a new measurement. No data is stored if canceled or before completion of the third measurement.
The UA-1030T detects the pulse and inflates the cuff to a systolic pressure level automatically. If re-inflation occurs repeatedly, set a fixed pressure value to avoid it:
Use this method when re-inflation occurs repeatedly even at 240 mmHg, or when results aren't displayed even as pressure decreases to 20 mmHg or less.
The monitor provides a blood pressure and pulse reading even when an irregular heartbeat occurs. The Irregular Heartbeat symbol appears if a heartbeat varies by 25% from the average of all heartbeat intervals during measurement. Relax, remain still, and refrain from talking during measurement. Contact your physician if you see this symbol frequently.
Provides a snapshot of your blood pressure classification based on your measurement, per the Seventh Report of the Joint National Committee (JNC7) on Prevention, Evaluation and Treatment of High Blood Pressure (National Heart, Lung and Blood Institute, May 2003). Because other risk factors (diabetes, obesity, smoking, etc.) affect blood pressure, the indicator is approximate — consult your physician for interpretation. Residents outside the U.S. (Canada, Mexico) should refer to the WHO Classification Table instead.
The device stores the last 90 measurements and announces the data values as they are displayed.
An AC adapter jack allows power from a home outlet. Use only the exclusive TB-233 AC adapter to avoid potential damage; contact A&D Medical (1-888-726-9966) or AutoControl in Canada (1-800-461-0991) to purchase one.
Note: If the monitor has batteries and an AC adapter connected, measurements stored in memory may be lost if the AC adapter is unplugged from the wall first. Ensure the unit is off and the monitor-side plug is removed quickly before unplugging from the wall.
Blood pressure is the force exerted by blood against artery walls. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. It is affected by age, weight, time of day, activity level, climate, altitude, season, caffeine/alcohol, medications, emotional stress, and tight clothing. It varies day to day and even rises during work/play, falling to its lowest during sleep.
| BP Classification | Systolic (mmHg) | Diastolic (mmHg) |
|---|---|---|
| Normal | less than 120 | and less than 80 |
| Prehypertension | 120–139 | or 80–89 |
| Stage 1 Hypertension | 140–159 | or 90–99 |
| Stage 2 Hypertension | ≥160 | or ≥100 |
Source: The Seventh Report of the Joint National Committee on Prevention, Evaluation and Treatment of High Blood Pressure. National Heart, Lung and Blood Institute — May 2003.
| BP Classification | Systolic (mmHg) | Diastolic (mmHg) |
|---|---|---|
| Optimal | less than 120 | and less than 80 |
| Normal | less than 130 | or less than 85 |
| High-Normal | 130–139 | or 85–89 |
| Grade 1 Hypertension | 140–159 | or 90–99 |
| Grade 2 Hypertension | 160–179 | or 100–109 |
Consistency is the most important method for getting an accurate blood pressure measurement. Devote two weeks to consistent monitoring — same time of day, location, chair, and cuff — to build a baseline you can compare future readings against.
Additional tools are available at www.andmedical.com: Large Print Instruction Manuals, Animated Operating Instructions, and Additional Logbook Sheets.
| Problem | Probable Cause | Corrective Action |
|---|---|---|
| Nothing appears in the display after pressing START | Batteries are drained; battery terminals not in the correct position | Replace all batteries with new alkaline batteries; reinstall matching negative/positive terminals |
| The cuff does not inflate | Battery voltage too low (low battery symbol blinks, or no symbol if fully drained) | Replace all batteries with new alkaline batteries |
| The unit does not measure; readings are too high or too low | The cuff is not fastened properly | Fasten the cuff correctly |
| You moved your arm or body during measurement | Remain very still and quiet during measurement | |
| The cuff position is not correct | Sit comfortably and still; make sure the cuff is level with your heart | |
| You are using the wrong size cuff | See "Selecting the Correct Cuff Size" | |
| Taking too many readings on the same arm in a short period | Relax several minutes before each measurement | |
| Value differs from a reading measured at a clinic or doctor's office | The healthcare professional may be using a different sized cuff | See "Why Your Readings Are Different Between Home and the Doctor's Office" |
| Your measurement may be elevated by "white coat hypertension" | See "Why Measure Blood Pressure at Home" |
Note: If the actions above do not solve the problem, call 1-888-726-9966. Do not attempt to repair the device yourself.
| Model | UA-1030T / UA-1030TCN (Canada) |
| Type | Oscillometric |
| Display | Digital, character height — Pressure (20.0mm) / Pulse (11.5mm) displayed simultaneously |
| Memory | 90 readings |
| Measurement Range | Pressure: 0–299 mmHg; Systolic: 60–279 mmHg; Diastolic: 40–200 mmHg; Pulse: 40–180 bpm |
| Accuracy | Pressure: ±3 mmHg or ±2% (whichever is greater); Pulse: ±5% |
| Power Source | 4 x 1.5V batteries (AA, LR6, or R6P) or 120V AC adapter (TB-233) |
| Classification | Type BF |
A&D Medical's full product line includes Digital Blood Pressure Monitors, Manual Blood Pressure Monitors, Blood Pressure Cuffs, Stethoscopes, Personal Health Scales, Digital Thermometers, Activity Monitors, and Wireless Products.
This monitor and cuff carry different warranties in different countries of purchase.