August 31, 2026 4 min read
Complete instruction manual for the A&D Medical / LifeSource UT-302 Instant-Read Ear Thermometer, covering setup, taking a reading, care, battery replacement, specs, and warranty.
Complete instruction manual for the A&D Medical / LifeSource® UT-302 Instant-Read Ear Thermometer, covering setup, taking a temperature reading, care, battery replacement, specifications, and warranty.
Package Contents: Thermometer body, 1x CR2032 lithium battery (pre-installed), instruction manual, fronthead cover.
The UT-302 uses infrared thermo-scanning to measure eardrum temperature. Because the eardrum shares blood vessels with the hypothalamus (the brain's temperature-control center), ear temperature is a reliable indicator of core body temperature. The unit scans the eardrum 512 times per second and displays the highest temperature detected — faster, safer, and easier than a traditional stick thermometer, and usable even on a sleeping child.
Press the Power/Scan button once, then press and hold it while pressing the “M” button to toggle between Celsius and Fahrenheit modes.
Q: Why do I get consistently low temperature readings?
A: The thermometer isn't seated properly — the tip must form a seal around the ear opening. Or the lens is dirty; clean with an alcohol-moistened cotton swab.
Q: Why do left-ear and right-ear readings differ?
A: Core temperature is roughly the same in both ears, but shape, cleanliness, and thermometer position can cause small differences between readings.
Q: Can I use this on an animal?
A: No — the UT-302 is programmed specifically for human ear anatomy and body temperature.
| Specification | Details |
|---|---|
| Model | UT-302 |
| Dimensions | 2.2" x 1.75" x 6.2" (55.3 x 44.4 x 158.5mm) |
| Weight | 2.3 oz. (65.5 g) |
| Sensor type | Thermopile Sensor ST118 |
| Measuring range | 50°F–122°F (10°C–50°C) |
| Measuring accuracy | ±0.4°F (0.2°C) from 96.8°F–102.2°F (36°C–39°C); ±0.5°F (0.3°C) outside that range |
| Display resolution | 0.1°F or °C |
| Response time | 1.0 second |
| Memory function | 10 readings |
| Automatic shutdown | After 1 minute of non-use |
| Power supply | One 3V CR-2032 lithium battery |
| Battery life | Approximately 5,000 measurements |
| Operating environment | 60.8°F–95°F (16°C–35°C), up to 95% relative humidity |
| Storage environment | -4°F–122°F (-20°C–50°C), up to 95% relative humidity |
The UT-302 is warranted to the original retail purchaser to be free of defects in material and workmanship for 2 years from date of purchase, excluding damage from misuse, abuse, alteration, improper components, or failure to follow operating instructions. Register online at www.LifeSourceOnline.com.
USA: A&D Medical, 1756 Automation Parkway, San Jose, CA 95131. LifeSource Health Line (toll-free): 1-888-726-9966.
Canada: Auto Control Medical, 6695 Millcreek Drive, Unit 5, Mississauga, Ontario L5N 5R8. Toll-free: 1-800-461-0991.