Complete instruction manual for the A&D Digital Blood Pressure Monitor, Model UA-767PBT-C — a Continua-certified, Bluetooth-enabled arm cuff monitor.
Complete instruction manual for the A&D Digital Blood Pressure Monitor, Model UA-767PBT-C — a Continua-certified, Bluetooth-enabled arm cuff monitor.
Caution: Insert batteries correctly or the device will not work. Do not mix old and new batteries. Remove batteries if the device won't be used for a long time, as they may leak.
Insert the air connector plug firmly into the air socket.
Model UA-767PBT-C is a PBT-C series device designed to connect only with Continua Managers.
If your systolic pressure is expected to exceed 230 mmHg, or you use the optional small cuff:
The UA-767PBT-C still provides a blood pressure and pulse reading even when an irregular heartbeat occurs. An irregular heartbeat is defined as one that varies by 25% from the average of all heartbeats during measurement. Remain relaxed, still, and silent during measurement. Contact your physician if this indicator appears frequently.
Measurement results transmit automatically to the paired Manager after each reading. If transmission fails, the device stores up to the last 25 measurements in memory and sends all stored data at the next successful connection, then clears memory.
| Problem | Possible Reason | Recommended Action |
|---|---|---|
| Nothing appears in the display | Batteries drained, or terminals reversed | Replace all batteries; reinstall matching polarity markings |
| The cuff does not inflate | Low battery power, or cuff not fastened properly | Replace batteries; fasten cuff correctly |
| The unit does not measure | Arm/body moved during measurement | Sit comfortably and remain still and quiet |
| Readings too high or too low | Cuff position incorrect, weak/irregular heartbeat, or comparing to a clinic reading | Raise arm so cuff is level with heart; see "Why measure at home" |
| Other issues | — | Remove batteries, reinsert properly, retry. Contact dealer if unresolved — do not open/repair the device (voids warranty) |
Do not open the device — it contains delicate electrical components and an intricate air unit. European customers: routine inspection every 2 years is recommended for regulatory compliance; not required for non-European customers. Contact your authorized dealer or A&D for maintenance.
| Type | UA-767PBT-C |
| Measurement method | Oscillometric |
| Measurement range | Pressure: 20–280 mmHg; Pulse: 40–200 beats/min |
| Accuracy | Pressure: ±3 mmHg or 2% (whichever greater); Pulse: ±5% |
| Power supply | 4 x 1.5V batteries (LR6, R6P, or AA) |
| Classification | Type BF |
| Clinical test | ANSI / AAMI SP-10 1987 |
| EMC | IEC 60601-1-2: 2001 |
| Wireless | WML-40AH (Mitsumi); Bluetooth Ver.2.1 Class 1 HDP; Continua certified |
| Operating condition | +10°C to +40°C / 30%–85% RH |
| Storage condition | −10°C to +60°C / 30%–85% RH |
| Dimensions | Approx. 147 (W) x 64 (H) x 110 (D) mm |
| Weight | Approx. 300 g (excluding batteries) |
Note: EMC guidance tables (emissions, immunity, recommended separation distances) are provided in the full manual for regulatory/clinical use and are omitted here for brevity — see the original PDF for complete details.