August 31, 2026 8 min read
Full instruction guide for the LifeSource UA-767T Talking One-Step Auto-Inflation Digital Blood Pressure Monitor, including cuff sizing, measurement steps, troubleshooting, and specifications.
Instruction guide for the LifeSource UA-767T Talking One-Step Auto-Inflation Digital Blood Pressure Monitor (A&D Medical) — a home blood pressure and pulse monitor with voice announcement and headphone jack.
The UA-767T is an oscillometric blood pressure monitor designed for home use by individuals 18 years or older to measure systolic/diastolic blood pressure and pulse rate. It is not designed to measure blood pressure in people with common arrhythmias (such as premature atrial or ventricular beats, or atrial fibrillation), and is not intended for ambulatory use. The monitor announces readings aloud and includes a headphone jack for private listening.
Daily self-monitoring of blood pressure is an important way individuals can contribute to maintaining cardiovascular health and catching hypertension before it causes serious consequences. Self-measurement at home supplements — but does not replace — a physician's readings, and provides a more complete blood pressure history over time. Only a qualified medical practitioner should interpret blood pressure results; consult your physician with any questions about your readings.
Using the correct cuff size is important for an accurate reading — a cuff that's too large produces readings lower than actual blood pressure, and a cuff that's too small produces readings that are higher. Measure your upper arm circumference at the midpoint between shoulder and elbow with your arm relaxed at your side.
| Arm Size | Recommended Cuff Size | Replacement Cuff Model # |
|---|---|---|
| 5.1" – 7.9" (13–20cm) | Small | UA-279 |
| 7.5" – 12.2" (19–31cm) | Medium | UA-280 |
| 11.8" – 17.7" (30–45cm) | Large | UA-281 |
To attach the cuff, insert the air connector at the end of the cuff tube into the air socket on the left side of the monitor. If you change cuff size, you may need to adjust the exhaust velocity rate (no adjustment is needed switching between a medium and large cuff with a valve).
Install 4 type AA (1.5 volt) batteries and attach the cuff to the monitor before use.
Replace batteries when the "Low Battery" symbol appears on the display.
Pressure Switch Setting Guide:
| Usual Systolic | Set Switch To |
|---|---|
| up to 120 | 150 |
| 121–150 | 180 |
| 151–180 | 210 |
| 181–210 | 240 |
| 211+ | See note below |
Note: If your systolic reading is above 210, hold the START button until the switch reaches a value 30 mmHg above the systolic reading. Maximum cuff pressure is 320 mmHg; if reached, the cuff will automatically exhaust to zero. If you set a pressure that's too high, interrupt inflation by pressing START and reset the switch to the next lower setting.
You can use the monitor with headphones for privacy — the headphone jack is on the lower right side of the monitor. Plug the headphone connector into the jack; using headphones ensures only the person using the device hears the announcements.
Exhaust velocity is the rate at which cuff pressure deflates during measurement. It's preset at the factory and generally doesn't need adjustment, but the rate is dependent on cuff size, not arm size — if using a cuff other than the one shipped with the unit, you may need to adjust it.
If the exhaust velocity is not between 2 and 5 mmHg, adjust it with the unit powered off:
Once adjusted for a new cuff size, no further adjustment should be needed unless you change cuff sizes again.
Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. It's measured in millimeters of mercury (mmHg).
Blood pressure is affected by age, weight, time of day, activity level, climate, altitude, and season. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg; sleeping can decrease systolic pressure by as much as 10 mmHg. Taking repeated readings without a 5–10 minute interval between them, or without raising your arm to let blood flow back to the heart, can also affect results. Diet beverages with caffeine, alcohol, certain medications, emotional stress, and even tight-fitting clothes can make a difference.
| Category | Systolic (mmHg) | Diastolic (mmHg) |
|---|---|---|
| Optimal | <120 | <80 |
| Normal | <130 | <85 |
| High Normal | 130–139 | 85–89 |
| Hypertension Stage 1 | 140–159 | 90–99 |
| Hypertension Stage 2 | 160–179 | 100–109 |
| Hypertension Stage 3 | ≥180 | ≥110 |
Hypertension is diagnosed when readings are consistently above normal. Left untreated, it can lead to strokes, heart attacks, and other diseases, and is often called the "silent killer" because it doesn't always produce symptoms. It can usually be controlled through lifestyle changes (not smoking, reducing salt/fat intake, maintaining a healthy weight, exercising regularly), reduced stress, appropriate medication managed by a physician, and regular checkups.
LifeSource monitors use the "oscillometric method" — a noninvasive blood pressure determination technique. The cuff is first inflated until the artery is fully blocked; the monitor then takes measurements as the cuff deflates, examining the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat.
| Symbol | Condition/Cause | Recommended Action |
|---|---|---|
| MEASUREMENT IN PROGRESS | Appears during measurement; blinks and beeps when pulse is detected. | Remain still and refrain from talking; measurement in progress. |
| LOW BATTERY | Battery voltage is too low for the monitor to work properly. | Replace all batteries with new ones. |
| ERROR 1 | Appears when systolic and diastolic measurements are within 10 mmHg of each other. | Check for air leakage; make sure the cuff tube is properly connected and the exhaust rate is between 2–5 mmHg increments. |
| ERROR 2 | Appears if the pressure value is unstable due to movement during reading. | Try the measurement again and remain very still. |
| ERROR 3 | Appears when pressure value did not increase during cuff inflation: 1) cuff not fastened, or 2) air connector is backwards or not connected properly. | Make sure the cuff tube is properly connected to the monitor, the cuff is securely fastened, the exhaust rate is between 2–5 mmHg, and the air connector is inserted properly into the tube and monitor. |
If you hear an announcement of a measurement error, take the corrective actions above and re-measure.
| Model | UA-767T |
| Type | Oscillometric |
| Display | Digital, 16mm character height; pressure and pulse displayed simultaneously |
| Measurement Range | Pressure: 20–280 mmHg; Pulse: 40–200 pulses/minute |
| Accuracy | Pressure: ±3 mmHg or 2% (whichever is greater); Pulse: ±5% |
| Pressurization | Automatic, via microbomb |
| Depressurization | Constant-exhaust valve system |
| Power Source | 4 alkaline AA batteries (1.5V), not included |
| Battery Life | Approximately 4 months with one measurement daily |
| Operating Range | 50–104°F (10–40°C), <85% relative humidity |
| Storage Range | -4 to 158°F (-20 to 70°C), <95% relative humidity |
| Dimensions | Length: 4.2" (106mm); Width: 6.4" (163mm); Height: 2.8" (70mm) |
| Weight | 11.3 oz (321 g) without batteries |
Blood pressure measurements taken by the UA-767T are equivalent to those obtained by a trained observer using the auscultation method with cuff and stethoscope, within the limits prescribed by the American National Standards Institute for electronic or automatic sphygmomanometers.
The UA-767T is designed for home use by people 18 years or older, to measure systolic/diastolic pressure and pulse rate. It is not designed to measure blood pressure in people with common arrhythmias (e.g., premature atrial or ventricular beats, or atrial fibrillation), and is not intended for ambulatory use.
LifeSource, a division of A&D Engineering, Inc. — 1555 McCandless Drive, Milpitas, CA 95035, USA. Toll-free LifeSource Health Line: 1-888-726-9966. www.LifeSourceOnline.com
Canada: Auto Control Medical, 206A boul. Brunswick, Pointe-Claire, Quebec H9R 5P9. Toll-free: 1-800-461-0991
This LifeSource product is covered by a lifetime warranty. If you would like to receive a Braille or audio manual for Model 767T, call the LifeSource Health Line at 1-888-726-9966.