  • LifeSource UA-767T Talking Blood Pressure Monitor — Instruction Guide

    August 31, 2026 8 min read

    Full instruction guide for the LifeSource UA-767T Talking One-Step Auto-Inflation Digital Blood Pressure Monitor, including cuff sizing, measurement steps, troubleshooting, and specifications.

    Full text manual (for reference)

    Instruction guide for the LifeSource UA-767T Talking One-Step Auto-Inflation Digital Blood Pressure Monitor (A&D Medical) — a home blood pressure and pulse monitor with voice announcement and headphone jack.

    Overview

    The UA-767T is an oscillometric blood pressure monitor designed for home use by individuals 18 years or older to measure systolic/diastolic blood pressure and pulse rate. It is not designed to measure blood pressure in people with common arrhythmias (such as premature atrial or ventricular beats, or atrial fibrillation), and is not intended for ambulatory use. The monitor announces readings aloud and includes a headphone jack for private listening.

    Daily self-monitoring of blood pressure is an important way individuals can contribute to maintaining cardiovascular health and catching hypertension before it causes serious consequences. Self-measurement at home supplements — but does not replace — a physician's readings, and provides a more complete blood pressure history over time. Only a qualified medical practitioner should interpret blood pressure results; consult your physician with any questions about your readings.

    Selecting the Correct Cuff

    Using the correct cuff size is important for an accurate reading — a cuff that's too large produces readings lower than actual blood pressure, and a cuff that's too small produces readings that are higher. Measure your upper arm circumference at the midpoint between shoulder and elbow with your arm relaxed at your side.

    Arm Size Recommended Cuff Size Replacement Cuff Model #
    5.1" – 7.9" (13–20cm) Small UA-279
    7.5" – 12.2" (19–31cm) Medium UA-280
    11.8" – 17.7" (30–45cm) Large UA-281

    To attach the cuff, insert the air connector at the end of the cuff tube into the air socket on the left side of the monitor. If you change cuff size, you may need to adjust the exhaust velocity rate (no adjustment is needed switching between a medium and large cuff with a valve).

    Before You Start — Installing Batteries

    Install 4 type AA (1.5 volt) batteries and attach the cuff to the monitor before use.

    1. Remove the battery compartment cover by gently pushing down on the arrow and sliding the cover forward.
    2. Put in the top row of batteries first, matching the positive (+) and negative (–) terminals to those indicated in the compartment. Make sure the batteries contact the compartment terminals.
    3. Replace the cover by sliding it into the compartment and gently pressing into place.

    Replace batteries when the "Low Battery" symbol appears on the display.

    Taking Your Blood Pressure

    Tips for Blood Pressure Monitoring

    • Relax for about 5 to 10 minutes before measurement.
    • Remove constricting clothing and place the cuff on a bare arm.
    • Unless your physician recommends otherwise, use your left arm.
    • Measure at the same time each day for a consistent history.

    Steps

    1. Sit comfortably with your left arm resting on a flat surface so the center of your upper arm is level with your heart.
    2. Lay your left arm on the table, palm up, and thread the cuff end through the metal loop, smooth side against the arm. Position the tube off-center toward the inner side of the arm, in line with the little finger.
    3. Pull the end of the cuff to tighten it, fold back the extra material, and fasten the Velcro. The cuff should be snug but not too tight — you should be able to insert two fingers between the cuff and your arm.
    4. Before measuring, the pressure switch should be set at least 30 mmHg above the expected systolic pressure — use the guide table below. If the monitor can't detect systolic pressure at the current setting, it will automatically re-inflate to the next higher setting. Don't worry if the switch is set too low.
    5. Press the START button, remain still, and refrain from talking while the cuff inflates to the correct level.
    6. It's normal for the cuff to feel very tight. All display symbols appear briefly, then blinking numerals starting at zero appear as the cuff pressurizes. When inflation completes, the MEASUREMENT IN PROGRESS symbol blinks and beeps with each pulse beat. (Press START at any time to stop inflation.)
    7. Watch the display as the cuff deflates — numbers should decline in increments of 2 to 5 mmHg. Wait for the long beep indicating measurement is complete; systolic/diastolic pressure and pulse rate are then displayed and announced before the cuff deflates.
    8. Example announcement for a systolic reading of 120, diastolic of 73, and pulse of 69: "Your blood pressure is 120 over 73 and pulse is 69." The announcement repeats four times before the monitor automatically shuts off. You can stop the announcement and turn off the monitor at any time by pressing START.
    9. Remove the cuff and record your blood pressure and pulse rate, including the date and time of measurement.

    Pressure Switch Setting Guide:

    Usual Systolic Set Switch To
    up to 120 150
    121–150 180
    151–180 210
    181–210 240
    211+ See note below

    Note: If your systolic reading is above 210, hold the START button until the switch reaches a value 30 mmHg above the systolic reading. Maximum cuff pressure is 320 mmHg; if reached, the cuff will automatically exhaust to zero. If you set a pressure that's too high, interrupt inflation by pressing START and reset the switch to the next lower setting.

    Using Headphones

    You can use the monitor with headphones for privacy — the headphone jack is on the lower right side of the monitor. Plug the headphone connector into the jack; using headphones ensures only the person using the device hears the announcements.

    Checking & Adjusting Exhaust Velocity

    Exhaust velocity is the rate at which cuff pressure deflates during measurement. It's preset at the factory and generally doesn't need adjustment, but the rate is dependent on cuff size, not arm size — if using a cuff other than the one shipped with the unit, you may need to adjust it.

    1. Put on the arm cuff.
    2. Press START. The unit begins taking a reading.
    3. Watch the display as the cuff deflates. When the MEASUREMENT IN PROGRESS symbol appears, numbers should decline in increments of 2 to 5 mmHg.

    If the exhaust velocity is not between 2 and 5 mmHg, adjust it with the unit powered off:

    1. With a fingernail, remove the thin black plastic disc covering the adjustment screw (skip if not present).
    2. With a small screwdriver, turn the adjustment screw approximately 45° to the right to increase exhaust velocity, or to the left to decrease it.
    3. Repeat the test until the exhaust velocity is between 2 and 5 mmHg.

    Once adjusted for a new cuff size, no further adjustment should be needed unless you change cuff sizes again.

    About Blood Pressure

    Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. It's measured in millimeters of mercury (mmHg).

    What Affects Blood Pressure

    Blood pressure is affected by age, weight, time of day, activity level, climate, altitude, and season. Walking can raise systolic pressure by 12 mmHg and diastolic pressure by 5.5 mmHg; sleeping can decrease systolic pressure by as much as 10 mmHg. Taking repeated readings without a 5–10 minute interval between them, or without raising your arm to let blood flow back to the heart, can also affect results. Diet beverages with caffeine, alcohol, certain medications, emotional stress, and even tight-fitting clothes can make a difference.

    Blood Pressure Evaluation (NIH JNC-VI Guidelines)

    Category Systolic (mmHg) Diastolic (mmHg)
    Optimal <120 <80
    Normal <130 <85
    High Normal 130–139 85–89
    Hypertension Stage 1 140–159 90–99
    Hypertension Stage 2 160–179 100–109
    Hypertension Stage 3 ≥180 ≥110

    Hypertension is diagnosed when readings are consistently above normal. Left untreated, it can lead to strokes, heart attacks, and other diseases, and is often called the "silent killer" because it doesn't always produce symptoms. It can usually be controlled through lifestyle changes (not smoking, reducing salt/fat intake, maintaining a healthy weight, exercising regularly), reduced stress, appropriate medication managed by a physician, and regular checkups.

    How the UA-767T Works

    LifeSource monitors use the "oscillometric method" — a noninvasive blood pressure determination technique. The cuff is first inflated until the artery is fully blocked; the monitor then takes measurements as the cuff deflates, examining the pulsatile pressure generated by the arterial wall as it expands and contracts against the cuff with each heartbeat.

    Display Symbols & Troubleshooting

    Symbol Condition/Cause Recommended Action
    MEASUREMENT IN PROGRESS Appears during measurement; blinks and beeps when pulse is detected. Remain still and refrain from talking; measurement in progress.
    LOW BATTERY Battery voltage is too low for the monitor to work properly. Replace all batteries with new ones.
    ERROR 1 Appears when systolic and diastolic measurements are within 10 mmHg of each other. Check for air leakage; make sure the cuff tube is properly connected and the exhaust rate is between 2–5 mmHg increments.
    ERROR 2 Appears if the pressure value is unstable due to movement during reading. Try the measurement again and remain very still.
    ERROR 3 Appears when pressure value did not increase during cuff inflation: 1) cuff not fastened, or 2) air connector is backwards or not connected properly. Make sure the cuff tube is properly connected to the monitor, the cuff is securely fastened, the exhaust rate is between 2–5 mmHg, and the air connector is inserted properly into the tube and monitor.

    If you hear an announcement of a measurement error, take the corrective actions above and re-measure.

    Specifications

    Model UA-767T
    Type Oscillometric
    Display Digital, 16mm character height; pressure and pulse displayed simultaneously
    Measurement Range Pressure: 20–280 mmHg; Pulse: 40–200 pulses/minute
    Accuracy Pressure: ±3 mmHg or 2% (whichever is greater); Pulse: ±5%
    Pressurization Automatic, via microbomb
    Depressurization Constant-exhaust valve system
    Power Source 4 alkaline AA batteries (1.5V), not included
    Battery Life Approximately 4 months with one measurement daily
    Operating Range 50–104°F (10–40°C), <85% relative humidity
    Storage Range -4 to 158°F (-20 to 70°C), <95% relative humidity
    Dimensions Length: 4.2" (106mm); Width: 6.4" (163mm); Height: 2.8" (70mm)
    Weight 11.3 oz (321 g) without batteries

    Blood pressure measurements taken by the UA-767T are equivalent to those obtained by a trained observer using the auscultation method with cuff and stethoscope, within the limits prescribed by the American National Standards Institute for electronic or automatic sphygmomanometers.

    Important Safety Information

    • Only a medical practitioner is qualified to interpret blood pressure results. This device should not replace regular medical checkups.
    • Your physician should observe how you use this device and should verify your blood pressure before prescribing or changing medication.
    • Consult your physician if you have any doubts about the results obtained. If mechanical problems occur, contact LifeSource.
    • Do not attempt to repair, calibrate, or modify this device.
    • Because the UA-767T contains delicate, high-precision parts, avoid exposing it to extreme temperature or humidity, direct sunlight, impacts, or dust. LifeSource guarantees accuracy only when the device is stored and used within the specified temperature/humidity ranges.
    • Clean the monitor and cuff with a soft, dry cloth, or one dampened with water and mild detergent. Never use alcohol, benzene, thinner, or other strong chemicals.
    • Remove batteries if the monitor will be unused for more than six months.

    Precautions

    The UA-767T is designed for home use by people 18 years or older, to measure systolic/diastolic pressure and pulse rate. It is not designed to measure blood pressure in people with common arrhythmias (e.g., premature atrial or ventricular beats, or atrial fibrillation), and is not intended for ambulatory use.

    Contact & Support

    LifeSource, a division of A&D Engineering, Inc. — 1555 McCandless Drive, Milpitas, CA 95035, USA. Toll-free LifeSource Health Line: 1-888-726-9966. www.LifeSourceOnline.com

    Canada: Auto Control Medical, 206A boul. Brunswick, Pointe-Claire, Quebec H9R 5P9. Toll-free: 1-800-461-0991

    This LifeSource product is covered by a lifetime warranty. If you would like to receive a Braille or audio manual for Model 767T, call the LifeSource Health Line at 1-888-726-9966.